Frequently Asked Questions about HdrOzone®
Yes
At least 5 paired cycles.
The cartridge is 120 ml and contains 59% H2O2.
They are supplied by the manufacturer or an authorized dealer.
Yes
Fabrics, cellulosic materials (textiles), silver-containing materials, liquids, powders, copper, natural rubber, and similar materials are not suitable for this sterilization method.
When power is restored, the device displays "OK?" / "Continue?" on the screen. Based on the user's selection, it will either resume the cycle or terminate it.
A connection to the hospital's central oxygen supply is required.
When used from the central oxygen system it should be greater than 80%; when used from a medical oxygen cylinder it should be greater than 98%.
Maintenance should be performed approximately every 6 months (every 300 cycles).
Yes, there should be an exhaust connection to vent spent ozone outside the room.
In HdrOzone mode the entire sterilization chamber can be fully loaded. (Max. 131 L)
Yes
Tyvek is used.
Use the chemical and biological indicators intended for VH2O2.
The device detects moist material, dries it, and continues sterilization without interruption.
No.
There are no limitations on lumen, diameter, or length. On a full load, all items can be sterilized together in mixed arrangement.
Yes. A minimum of 100 cycle records are retained.
Yes. The device also includes quarter-cycle and half-cycle validation programs.
Because full-load and mixed-load sterilization can be completed in a single cycle, it is approximately one-third the cost.
No.
No.
No.
No. The resulting end products are water vapor and oxygen.
No. It can be installed in any area where a hydrogen peroxide device is installed.